Calibration of Big Bottle RAD H2O set-up for radon in water using HDPE bottles
Language
English
Obiettivo Specifico
2IT. Laboratori analitici e sperimentali
Status
Published
JCR Journal
JCR Journal
Peer review journal
Yes
Journal
Issue/vol(year)
/76 (2015)
Pages (printed)
1-7
Date Issued
March 17, 2015
Abstract
Glass bottles are generally employed for water sampling because glass is impervious to radon and is not
lost during sample storage. On the other hand, glass is fragile and may break, so 1 L High Density
PolyEthylene (HDPE) bottles (Thermo Scientific Nalgene) are tested in place of glass vessels employing
Big Bottle RAD H2O device (Durridge Company) coupled with RAD7 monitor. The purpose of this calibration
is to quantify radon loss during storage in HDPE bottles, evaluate possible radon uptake by
known volume of desiccant (Drierite, granular CaSO4) and quantify radon interaction with the rubber and
plastic parts of the experimental circuit. These processes have been separately investigated, performing
proper experiments for the assessment of their influence on resulting radon data using seven series of
solutions at known activity concentrations in the range from 27 to 194 Bq/L. Percent radon loss during
storage in 1 L HDPE bottles has been estimated at 0.0045 min-1. Radon absorption by desiccant,
expressed as ‘equivalent’ volume of Drierite is 0.673 ± 0.092 L and is somehow independent, within
errors, from i) the amount of water already absorbed in Drierite, ii) a recirculation time greater than
30 min and iii) radon concentrations. Radon absorption/desorption from rubber and plastic parts of the
experimental device has been assessed as a function of concentration gradient between the inner volume
of the circuit and the pores of polymer's. A final algorithm accounting for the above described physical
processes has been developed for long runs (2-3 h). A simplified calculation method for short measurements
(30 min) is also provided.
lost during sample storage. On the other hand, glass is fragile and may break, so 1 L High Density
PolyEthylene (HDPE) bottles (Thermo Scientific Nalgene) are tested in place of glass vessels employing
Big Bottle RAD H2O device (Durridge Company) coupled with RAD7 monitor. The purpose of this calibration
is to quantify radon loss during storage in HDPE bottles, evaluate possible radon uptake by
known volume of desiccant (Drierite, granular CaSO4) and quantify radon interaction with the rubber and
plastic parts of the experimental circuit. These processes have been separately investigated, performing
proper experiments for the assessment of their influence on resulting radon data using seven series of
solutions at known activity concentrations in the range from 27 to 194 Bq/L. Percent radon loss during
storage in 1 L HDPE bottles has been estimated at 0.0045 min-1. Radon absorption by desiccant,
expressed as ‘equivalent’ volume of Drierite is 0.673 ± 0.092 L and is somehow independent, within
errors, from i) the amount of water already absorbed in Drierite, ii) a recirculation time greater than
30 min and iii) radon concentrations. Radon absorption/desorption from rubber and plastic parts of the
experimental device has been assessed as a function of concentration gradient between the inner volume
of the circuit and the pores of polymer's. A final algorithm accounting for the above described physical
processes has been developed for long runs (2-3 h). A simplified calculation method for short measurements
(30 min) is also provided.
References
Arafa, W., 2002. Permeability of radon222 through some materials. Radiat. Meas. 35, 207e211.
Ashry, A.H., Abou-Leila, M., Abdalla, A.M., 2011. Measurement of radon permeability through polyethylene membrane using scintillation detector. Radiat. Meas. 46, 149-152.
Fernandez, P.L., Quindos, L.S., Sainz, C., Gomez, J., 2004. A theoretical approach to the measurement of radon diffusion and adsorption coefficients in radonproof membranes. Nucl. Instrum. Methods Phys. Res. B 217, 167-176.
Galli, G., Guadoni, C., Mancini, C., 1999. Radon grab sampling in water by means of radon transfer in activated charcoal collectors. In: Proceedings of the Fourth International Conference on Rare Gases Geochemistry. 8e10 October, 1997, vol. 22C. University of ROMA TRE e Italy. Il Nuovo Cimento, pp. 583-587, 3-4.
Jiranek, M., Rovenska, K., Fronka, A., 2008. Radon diffusion coefficient e a material property determining the applicability of waterproof membranes as radon barriers. In: Proceedings of the American Association of Radon Scientists and Technologists 2008 International Symposium, Las Vegas NV, September 14-17, 2008.
Leaney, F.W., Herczeg, A., 2006. A rapid field extraction method for determination of radon-222 in natural waters by liquid scintillation counting. Limnol. Oceanogr. Methods 4, 254-259.
Saito, M., 1983. Correction for loss of radon-222 in water sample caused by the use of a polyethylene bottle. Radioisotopes 32, 109-112.
Weigel, F., 1978. Radon. Chem. Ztg. 102, 287-299.
Ashry, A.H., Abou-Leila, M., Abdalla, A.M., 2011. Measurement of radon permeability through polyethylene membrane using scintillation detector. Radiat. Meas. 46, 149-152.
Fernandez, P.L., Quindos, L.S., Sainz, C., Gomez, J., 2004. A theoretical approach to the measurement of radon diffusion and adsorption coefficients in radonproof membranes. Nucl. Instrum. Methods Phys. Res. B 217, 167-176.
Galli, G., Guadoni, C., Mancini, C., 1999. Radon grab sampling in water by means of radon transfer in activated charcoal collectors. In: Proceedings of the Fourth International Conference on Rare Gases Geochemistry. 8e10 October, 1997, vol. 22C. University of ROMA TRE e Italy. Il Nuovo Cimento, pp. 583-587, 3-4.
Jiranek, M., Rovenska, K., Fronka, A., 2008. Radon diffusion coefficient e a material property determining the applicability of waterproof membranes as radon barriers. In: Proceedings of the American Association of Radon Scientists and Technologists 2008 International Symposium, Las Vegas NV, September 14-17, 2008.
Leaney, F.W., Herczeg, A., 2006. A rapid field extraction method for determination of radon-222 in natural waters by liquid scintillation counting. Limnol. Oceanogr. Methods 4, 254-259.
Saito, M., 1983. Correction for loss of radon-222 in water sample caused by the use of a polyethylene bottle. Radioisotopes 32, 109-112.
Weigel, F., 1978. Radon. Chem. Ztg. 102, 287-299.
Type
article
File(s)![]()
![]()
Name
article.pdf
Size
969.99 KB
Format
Adobe PDF
Checksum (MD5)
The MD5 message-digest algorithm can be used to verify the integrity of the file that you have uploaded. You can calculate its value locally via tools that are generally available in each operative system like md5sum
a7e8371040c9040a58552d60e318c640
Name
draft.pdf
Size
506.87 KB
Format
Adobe PDF
Checksum (MD5)
The MD5 message-digest algorithm can be used to verify the integrity of the file that you have uploaded. You can calculate its value locally via tools that are generally available in each operative system like md5sum
4567ec29107ad9d3a617fd3c46f8eee5