Diode laser-based gas analyser for the simultaneous measurement of CO2 and HF in volcanic plumes
Language
English
Obiettivo Specifico
5V. Processi eruttivi e post-eruttivi
Status
Published
JCR Journal
JCR Journal
Issue/vol(year)
/11(2018)
Pages (printed)
329–339
Date Issued
2018
Abstract
A portable analyser is described for simultane- ous detection of CO2 and HF emitted by volcanoes and fumaroles. The system is based on two fibre-coupled dis- tributed feedback lasers and only one multipass cell, and provides the absolute concentration values of the two gases, without requiring a calibration procedure, at a maximum rate of 4 Hz. The spectrometer can operate both in a closed-cell configuration and in an open-cell set-up, with the latter mit- igating problems associated with chemisorption of the HF molecule. The concept, practical realization and laboratory performance of the device are presented, together with results from a first test campaign measuring volcanic gases emitted from the crater of Vulcano, Italy. We obtained an in-field sen- sitivity of 320 ppb for CO2 and 20 ppb for HF at 2 s integration time.
Type
article
File(s)![]()
Name
amt-11-329-2018.pdf
Size
6.28 MB
Format
Adobe PDF
Checksum (MD5)
The MD5 message-digest algorithm can be used to verify the integrity of the file that you have uploaded. You can calculate its value locally via tools that are generally available in each operative system like md5sum
bc5f78ab43fed79895e1ac058be3269f