Earth-prints repository, logo   DSpace

About DSpace Software
|earth-prints home page | roma library | bologna library | catania library | milano library | napoli library | palermo library
Please use this identifier to cite or link to this item: http://hdl.handle.net/2122/7070

Authors: Cusano, P.*
Bobbio, A.*
Festa, G.*
Petrosino, S.*
Rombetto, S.*
Title: Analysis of seismic noise to check the mechanical isolation
Other Titles: MECHANICAL DISTURBANCE ON A MEDICAL DEVICE
Title of journal: Annals of geophysics
Series/Report no.: 3/54 (2011)
Publisher: Istituto Nazionale di Geofisica e Vulcanologia
Issue Date: 6-Jul-2011
DOI: 10.4401/ag-4774
URL: http://www.annalsofgeophysics.eu/index.php/annals/article/view/4774
Keywords: Seismic noise
Spectral ratio analyses
Magnetoencephalography
Abstract: We have investigated the mechanical response of a magnetically shielded room that hosts a magnetoencephalography system that is subject to external vibrations. This is a superconducting quantum interference device, which are the most sensitive sensors for magnetic flux variations. When the magnetoencephalography operates with people inside the room, the spectrum of the flux of the magnetic field shows anomalous peaks at several frequencies between 1 Hz and 20 Hz, independent of the experiment that is being run. As the variations in the flux of the magnetic field through the sensors might not only be related to the electrical currents circulating inside the brain, but also to non-damped mechanical oscillations of the room, we installed seismic instrumentation to measure the effective motion inside the room and to compare it to the external motion. For this analysis, we recorded the ambient seismic noise at two very close stations, one inside the magnetically shielded room, the other one outside in the room in which the magnetically shielded room is itself located. Data were collected over four days, including a week-end, to study the response of the magnetically shielded room subjected to different energy levels of external vibrations. The root mean square, Fourier spectra and power spectral density show significant differences between the signal recorded inside and outside the magnetically shielded room, with several anomalous peaks in the frequency band of 1 Hz to 20 Hz. The normalized spectral quantities (horizontal to vertical spectral ratio, and ratio between the internal and external spectra) show large amplification at several frequencies, reaching in some cases one order of magnitude. We concluded that the magnetically shielded room does not dampen the external vibrations, but it instead appears to amplify these across a broad frequency range.
Appears in Collections:04.06.04. Ground motion
Papers Published / Papers in press

Files in This Item:

File Description SizeFormatVisibility
CusanoEtAl2011AG.pdfArticle1.3 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.


Share this record
Del.icio.us

Citeulike

Connotea

Facebook

Stumble it!

reddit


 

Valid XHTML 1.0! ICT Support, development & maintenance are provided by CINECA. Powered on DSpace Software. CINECA