Please use this identifier to cite or link to this item: http://hdl.handle.net/2122/4307
AuthorsAchilli, V.* 
Baldi, P.* 
Casula, G.* 
Errani, M.* 
Focardi, S.* 
Guerzoni, M.* 
Palmonari, F.* 
Ragunì, G.* 
TitleA calibration system for superconducting Gravimeters
Issue DateOct-1994
Series/Report no.69 /(1995)
URIhttp://hdl.handle.net/2122/4307
KeywordsGravimetry
Super Conducting Gravimeters
Calibration
Subject Classification01. Atmosphere::01.01. Atmosphere::01.01.08. Instruments and techniques 
AbstractA new method for the calibration of a superconducting gravimeter is described, in wihich a 273 Kg annular mass is placed around the meter and is moved up and down. The geometry of the apparatus is easy to model and the accuracy in the computation of the gravity variation induced by the mass, 6.7 microGal, is limited only by the accuracy in the knowledge of value of the gravitational constant. Measurements done in '91 and '92 for the calibration of the instrument GRW-T015 are described. The calibration factor has been determined with a precision of about 0.3%.